Siemens Healthcare Diagnostics Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge recall — FDA Z-2450-2015

customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In…

Class IIclassification
Z-2450-2015official record ID
Sep 2, 2015report date

Recall record at a glance

Official recordZ-2450-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Healthcare Diagnostics, Inc.
Report dateSep 2, 2015
Recall initiation dateJul 23, 2015
QuantityTotal 1655 units
LocationNewark, DE, United States

Official product description

Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B¿2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B¿2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Why the product was recalled

customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.

Distribution information

US Nationwide Distribution in the states of TN, AZ, CT, Fl, CA, NY and WI

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.