Siemens Healthcare Diagnostics Dimension Vista 500, model no. 10284473 an in vitro recall — FDA Z-0616-2019

Siemens Healthcare Diagnostics has identified an issue with Dimension Vista software v.3.7 and v.3.8.1 that may result in aliquot well contamination. If a…

Class IIclassification
Z-0616-2019official record ID
Dec 26, 2018report date

Recall record at a glance

Official recordZ-0616-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Healthcare Diagnostics, Inc.
Report dateDec 26, 2018
Recall initiation dateOct 16, 2018
Quantity1649
LocationNewark, DE, United States

Official product description

Dimension Vista 500, model no. 10284473 an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.

Why the product was recalled

Siemens Healthcare Diagnostics has identified an issue with Dimension Vista software v.3.7 and v.3.8.1 that may result in aliquot well contamination. If a test is cancelled by the operator while the instrument is aliquotting the sample, the Dimension Vista instrument may continue to deliver the cancelled sample or vial contents to the aliquot well(s). Future samples may be assigned to the same aliquot well(s) and become contaminated with the existing contents or overflow of the aliquot well may occur potentially contaminating a surrounding well(s). This issue only occurs when an operator manually cancels a test(s) for calibration, quality control (QC) or patient sample while the instrument is in the process of aliquotting the sample.

Distribution information

Nationwide/worldwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.