Recall record at a glance
| Official record | Z-2294-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Shimadzu Medical Systems |
| Report date | Jun 10, 2026 |
| Recall initiation date | Apr 20, 2026 |
| Quantity | 3487 units |
| Location | Long Beach, CA, United States |
Official product description
Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 / D150GC-40; Digital Radiography System RADspeed Pro: UD150B-40 / UD150V-40 / UD150L-40 / UD150L-40E/D150BC-40 / D150VC-40 / D150LC-40 / D150LC-40E/D150BC-41 / D150VC-41 / D150LC-41; Digital Radiography System RADspeed safire: UD150B-40 / D150BC-40; Radiography System RADspeed fit: N/A (Integrated in the system); X-RAY R/F SYSTEM SONIALVISION safire17:UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION safire Plus: UD150B-40 / D150BC-4;0 X-RAY R/F SYSTEM SONIALVISION safire: UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION VERSA: UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM FLUOROspeed 300:UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM SONIALVISION G4:D150BC-40S; X-RAY R/F SYSTEM FLUOROspeed X1:D150BC-40S
Why the product was recalled
Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.
Distribution information
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, The Bahamas, China, Denmark, Haiti.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: not stated.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.