FDA recall record

Shimadzu Medical Systems Digital Angiography System Trinias:UD150G-40 / D150GC-40 recall — FDA Z-2294-2026

Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

Class IIclassification
Z-2294-2026official record ID
Jun 10, 2026report date

Recall record at a glance

Official recordZ-2294-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmShimadzu Medical Systems
Report dateJun 10, 2026
Recall initiation dateApr 20, 2026
Quantity3487 units
LocationLong Beach, CA, United States

Official product description

Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 / D150GC-40; Digital Radiography System RADspeed Pro: UD150B-40 / UD150V-40 / UD150L-40 / UD150L-40E/D150BC-40 / D150VC-40 / D150LC-40 / D150LC-40E/D150BC-41 / D150VC-41 / D150LC-41; Digital Radiography System RADspeed safire: UD150B-40 / D150BC-40; Radiography System RADspeed fit: N/A (Integrated in the system); X-RAY R/F SYSTEM SONIALVISION safire17:UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION safire Plus: UD150B-40 / D150BC-4;0 X-RAY R/F SYSTEM SONIALVISION safire: UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION VERSA: UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM FLUOROspeed 300:UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM SONIALVISION G4:D150BC-40S; X-RAY R/F SYSTEM FLUOROspeed X1:D150BC-40S

Why the product was recalled

Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

Distribution information

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, The Bahamas, China, Denmark, Haiti.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: not stated.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.