FDA recall record

DeVilbiss Healthcare DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q recall — FDA Z-0724-2023

The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with…

Class IIclassification
Z-0724-2023official record ID
Dec 28, 2022report date

Recall record at a glance

Official recordZ-0724-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDeVilbiss Healthcare LLC
Report dateDec 28, 2022
Recall initiation dateNov 18, 2022
Quantity1004315 (US); 414559 (OUS)
LocationSomerset, PA, United States

Official product description

DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q

Why the product was recalled

The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.

Distribution information

US Nationwide. Foreign distribution pending.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.