FDA recall record

Ethicon DERMABOND" PRINEO" Skin Closure System Product Usage: DERMABOND PRINEO System is recall — FDA Z-1147-2018

Ethicon discovered that specific lots of DERMABOND"PRINEO"System may not dry within the specified time after proper application, and thus may fall off.

Class IIclassification
Z-1147-2018official record ID
Mar 28, 2018report date

Recall record at a glance

Official recordZ-1147-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmEthicon, Inc.
Report dateMar 28, 2018
Recall initiation dateNov 27, 2017
Quantity20,090 eaches (10,545 units)
LocationSomerville, NJ, United States

Official product description

DERMABOND" PRINEO" Skin Closure System Product Usage: DERMABOND PRINEO System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND PRINEO System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound durinQ application of the liquid adhesive

Why the product was recalled

Ethicon discovered that specific lots of DERMABOND"PRINEO"System may not dry within the specified time after proper application, and thus may fall off.

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.