FDA recall record

Limacorporate S.p.A Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during recall — FDA Z-0068-2018

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to…

Class IIclassification
Z-0068-2018official record ID
Nov 15, 2017report date

Recall record at a glance

Official recordZ-0068-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusCompleted
Recalling firmLimacorporate S.p.A
Report dateNov 15, 2017
Recall initiation dateSep 8, 2017
Quantity20 units in (US)
LocationSan Daniele del Friuli, N/A, Italy

Official product description

Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System

Why the product was recalled

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

Distribution information

Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.