FDA recall record

Defibtech DDU-100 Lifeline Automated External Defibrillator (AED)-Indicated for use on recall — FDA Z-1312-2023

A component of a sub-assembly used in the affected AEDs had not undergone durability and reliability testing, and if a malfunction occurs, therefore the…

Class IIclassification
Z-1312-2023official record ID
Apr 5, 2023report date

Recall record at a glance

Official recordZ-1312-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDefibtech, LLC
Report dateApr 5, 2023
Recall initiation dateFeb 14, 2023
Quantity29 units
LocationGuilford, CT, United States

Official product description

DDU-100 Lifeline Automated External Defibrillator (AED)-Indicated for use on victims of sudden cardiac arrest

Why the product was recalled

A component of a sub-assembly used in the affected AEDs had not undergone durability and reliability testing, and if a malfunction occurs, therefore the device could fail to deliver therapy.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.