FDA recall record

Maquet Datascope Corp - Cardiac Assist Division Datascope System 98/98xt, CS 100/CS100i and recall — FDA Z-1523-2014

Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and…

Class Iclassification
Z-1523-2014official record ID
May 28, 2014report date

Recall record at a glance

Official recordZ-1523-2014
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmMaquet Datascope Corp - Cardiac Assist Division
Report dateMay 28, 2014
Recall initiation dateMar 21, 2014
Quantity11,881 units (US 5183 and OUS 6698)
LocationMahwah, NJ, United States

Official product description

Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.

Why the product was recalled

Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.

Distribution information

Worldwide Distribution - USA (nationwide) and approximately 104 countries.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.