FDA recall record

Siemens Healthcare Diagnostics Dade Actin Activated Cephaloplastin Reagent recall — FDA Z-1703-2015

Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been…

Class IIclassification
Z-1703-2015official record ID
Jun 10, 2015report date

Recall record at a glance

Official recordZ-1703-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Healthcare Diagnostics, Inc.
Report dateJun 10, 2015
Recall initiation dateMar 25, 2015
Quantity27501
LocationNewark, DE, United States

Official product description

Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determination of the activated partial thromboplastin time (APTT) and other coagulation procedures requiring an activated partial thromboplastin reagent.

Why the product was recalled

Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been observed. Patient samples with recoveries close to the medical decision points could exhibit a deviation up to 4 seconds for the normal range and up to 33% for the pathological range.

Distribution information

Worldwide Distribution. US naionwide including Puerto Rico, Barbados and Mexico.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records