FDA recall record

BC Group International DA-2006P - Defibrillator / Pacer Analyzer recall — FDA Z-0457-2016

Device is not functioning as intended: Two wires running to the ECG distribution circuit board are reversed resulting in inverted waveforms.

Class IIIclassification
Z-0457-2016official record ID
Dec 30, 2015report date

Recall record at a glance

Official recordZ-0457-2016
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmBC Group International Inc.
Report dateDec 30, 2015
Recall initiation dateOct 28, 2015
Quantity5 units
LocationSaint Charles, MO, United States

Official product description

DA-2006P - Defibrillator / Pacer Analyzer; Product Usage: Intended to measure the output of a defibrillator or transcutaneous pacemaker.

Why the product was recalled

Device is not functioning as intended: Two wires running to the ECG distribution circuit board are reversed resulting in inverted waveforms.

Distribution information

US Nationwide Distribution in the states of South Carolina, Florida, and Canada.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.