FDA recall record

Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile recall — FDA Z-1819-2013

On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm…

Class IIclassification
Z-1819-2013official record ID
Aug 7, 2013report date

Recall record at a glance

Official recordZ-1819-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCustom Spine, Inc.
Report dateAug 7, 2013
Recall initiation dateMar 28, 2013
Quantity18 units
LocationParsippany, NJ, United States

Official product description

Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus Dr. Parsippany, NJ 07054 USA. TEL: 973-808-0019 The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine for pedicular and non-pedicular fixation.

Why the product was recalled

On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm.

Distribution information

Nationwide Distribution including Nebraska and Virginia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.