American Contract Systems Custom procedural convenience kits and trays, general & plastic recall — FDA Z-0825-2024

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process.…

Class IIclassification
Z-0825-2024official record ID
Feb 7, 2024report date

Recall record at a glance

Official recordZ-0825-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAmerican Contract Systems, Inc.
Report dateFeb 7, 2024
Recall initiation dateNov 29, 2023
Quantity7881 units
LocationTemple Terrace, FL, United States

Official product description

Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48

Why the product was recalled

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.