FDA recall record

American Contract Systems Custom procedural convenience kits and trays, gastroenterological recall — FDA Z-0822-2024

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process.…

Class IIclassification
Z-0822-2024official record ID
Feb 7, 2024report date

Recall record at a glance

Official recordZ-0822-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAmerican Contract Systems, Inc.
Report dateFeb 7, 2024
Recall initiation dateNov 29, 2023
Quantity3196 units
LocationTemple Terrace, FL, United States

Official product description

Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit number ANPR34S; c) LAP CHOLE - 213388, kit number BFLC02AD; d) GENERAL LAPAROSCOPY PACK, kit number GUGE09I; e) LAP CHOLE PACK, kit number HNLC80I; f) Lap Chole Ovarian Appy, kit number JSLC06I; g) LAP CHOLE, ST LUKES, kit number LMLP08Z; h) ROBOTIC PACK - 242561, kit number MHRB88AJ; i) KIT, GENERAL LAPAROSCOPY, kit number MMLP40F; j) KIT, MAJOR LAPAROTOMY, kit number MMML26F; k) LAP CHOLE - LAP HERNIA PACK - 282807, kit number OWLC07K; l) GENERAL LAPAROSCOPY SAH, kit number SAGL57H; m) ROBOTIC PACK SAH, kit number SARB30K; n) ROBOTIC PACK SAH, kit number SARB30L; o) ROBOTIC PACK SAH, kit number SARB30M; p) LAP CHOLE PACK, kit number SMLC53; q) ROBOTICS PACK, kit number SNRP85D; r) ROBOTIC PACK, kit number SSRO22C; s) ROBOTIC PACK, kit number SSRO22D; t) TPK DAVINCI PROSTATE SYSTEM, kit number TPDV16; u) ASC LAP CHOLE PACK (PS 42279), kit number UICH53Y; v) GENERAL LAPAROSCOPIC PACK, kit number WVGL60E

Why the product was recalled

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.