Recall record at a glance
| Official record | Z-2409-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Immuno-Mycologics, Inc. |
| Report date | Jul 18, 2018 |
| Recall initiation date | Mar 30, 2018 |
| Quantity | 875 units |
| Location | Norman, OK, United States |
Official product description
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Model # CR2003 Product Usage: The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of the capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gattii) in serum and cerebral spinal fluid (CSF). The CrAg Lateral Flow Assay is a prescription-use laboratory assay which can aid in the diagnosis of cryptococcosis.
Why the product was recalled
Potential for LFA strips to improperly wick which may cause false negative cryptococcosis test results
Distribution information
On March 30, 2018, IMMY began issuing URGENT MEDICAL DEVICE RECALL notices to customers via email. Customers were advised on how to recognize that the device has failed. The following are Actions to be taken by the customer/user: " You may continue using the affected lots. Any testing performed on defective strips should be considered invalid and should be repeated. " Please immediately notify lab personnel who are running the test to look for defective strips (as seen by fast wicking and/or very weak control lines). " Please immediately complete the Acknowledgement and Receipt Form even if you do not have any affected stock remaining and submit via email to: customerservice@immy.com. " Ensure relevant staff members are informed of this recall, including relevant clinicians. " If you have supplied any potentially affected product to another organization, please advise that organization of this recall and notify IMMY. " In case product is in transit, display this letter in a prominent place for one month. Should you need a replacement kit, you will need to return your affected inventory. Instructions will be provided by customer service on how to return kits to IMMY. If you have any questions, contact IMMYs Quality Assurance Department by calling 1-405-360-4669 Monday through Friday 8:30 AM to 5:00 PM Central Standard Time.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.