FDA recall record

Belmont Instrument CritiCool, thermoregulation device recall — FDA Z-0153-2023

Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The issue may…

Class IIclassification
Z-0153-2023official record ID
Nov 9, 2022report date

Recall record at a glance

Official recordZ-0153-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBelmont Instrument Corporation
Report dateNov 9, 2022
Recall initiation dateJul 20, 2021
Quantity158 (US); 803 (OUS)
LocationBillerica, MA, United States

Official product description

CritiCool, thermoregulation device

Why the product was recalled

Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The issue may potentially lead to delayed treatment and a failure to deliver the prescribed thermoregulation therapy. To date, the firm has received one report of death potentially related to this issue.

Distribution information

Domestic distribution nationwide. Foreign distribution worldwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.