Recall record at a glance
| Official record | Z-1682-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Mentor Texas, LP. |
| Report date | Apr 8, 2026 |
| Recall initiation date | Feb 18, 2026 |
| Location | Irving, TX, United States |
Official product description
CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-157LH Smooth Low Height Tissue Expander, 650cc
Why the product was recalled
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Distribution information
Worldwide distribution - US Nationwide and the countries of Canada, Albania, Australia, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, Iceland, Ireland, Italy, Norway, Pakistan, Poland, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Czech Republic.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.