FDA recall record

Covidien Tri-Staple 2.0 Reinforced Intelligent Reload 60 mm Medium/Thick-For use with Signia recall — FDA Z-1419-2021

Potential for the pouch to contain an extra laminate layer, creating the potential for the device sterility to be compromised. The inclusion of an…

Class IIclassification
Z-1419-2021official record ID
Apr 21, 2021report date

Recall record at a glance

Official recordZ-1419-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCovidien LLC
Report dateApr 21, 2021
Recall initiation dateMar 23, 2021
Quantity222 units
LocationMansfield, MA, United States

Official product description

Covidien Tri-Staple 2.0 Reinforced Intelligent Reload 60 mm Medium/Thick-For use with Signia stapling system. Staple line reinforcement has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas. Product Number: SIGTRSB60AMT

Why the product was recalled

Potential for the pouch to contain an extra laminate layer, creating the potential for the device sterility to be compromised. The inclusion of an additional laminate layer in the package may also impact complete de-gassing of the sterilization gas, Ethylene Oxide and may result in a potentially increased risk for patient infection and/or potential exposure to Ethylene Oxide retained in the pouch

Distribution information

AL, AR, FL, IN, LA, MA , MN, MO, NJ, NY, OR, PA, TX

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.