FDA recall record

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long recall — FDA Z-1383-2021

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If…

Class IIclassification
Z-1383-2021official record ID
Apr 21, 2021report date

Recall record at a glance

Official recordZ-1383-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCovidien, LP
Report dateApr 21, 2021
Recall initiation dateMar 4, 2021
Quantity5928 units WW: 4212 US 1716 OUS
LocationNorth Haven, CT, United States

Official product description

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT

Why the product was recalled

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Distribution information

Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.