FDA recall record

Beckman Coulter Coulter PrepPlus, Part Number 286600, Product Usage: The COULTER PrepPlus is recall — FDA Z-2689-2017

When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the…

Class IIclassification
Z-2689-2017official record ID
Jul 12, 2017report date

Recall record at a glance

Official recordZ-2689-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBeckman Coulter Inc.
Report dateJul 12, 2017
Recall initiation dateMay 22, 2017
Quantity43 units
LocationBrea, CA, United States

Official product description

Coulter PrepPlus, Part Number 286600, Product Usage: The COULTER PrepPlus is a microprocessor-controlled pipetting and diluting system designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.

Why the product was recalled

When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the absolute counts on only the first tube in a run.

Distribution information

Worldwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.