FDA recall record

Zeltiq Aesthetics CoolSculpting Elite System The CoolSculpting System SW release 2.0, SW recall — FDA Z-2383-2021

An incorrect error messaging system that could potentially lead to: 1) Reporting a thermal event error causing a user to re-treating the affected anatomic…

Class IIclassification
Z-2383-2021official record ID
Sep 8, 2021report date

Recall record at a glance

Official recordZ-2383-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZeltiq Aesthetics, Inc.
Report dateSep 8, 2021
Recall initiation dateJul 5, 2021
Quantity911 systems (update 08/23/2021 = 1105 systems)
LocationPleasanton, CA, United States

Official product description

CoolSculpting Elite System The CoolSculpting System SW release 2.0, SW release 3.0, SW release 3.0.1, SW release 3.0.2

Why the product was recalled

An incorrect error messaging system that could potentially lead to: 1) Reporting a thermal event error causing a user to re-treating the affected anatomic area within 24 hours, 2) Not reporting a thermal event or any other error codes causing a user to continue treating without being aware that a thermal event has occurred.

Distribution information

U.S.: AL, AR, AZ, CA, CO, CT, FL, GA,HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Italy, Norway, Sweden, France, Netherlands, Germany, Spain, United Kingdom, Portugal, South Korea, Singapore, and Hong Kong.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.