FDA recall record

Centurion Medical Products Convenience Kits from Centurion that contain an In-Vision- Plus recall — FDA Z-0861-2016

Convenience Kits contain an In-Vision-Plus Needleless IV Connector, code RYM5001 that is being recalled by the manufacturer, RyMed Technologies, LLC. A…

Class IIclassification
Z-0861-2016official record ID
Mar 2, 2016report date

Recall record at a glance

Official recordZ-0861-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCenturion Medical Products Corporation
Report dateMar 2, 2016
Recall initiation dateJan 19, 2016
Quantity1130
LocationWilliamston, MI, United States

Official product description

Convenience Kits from Centurion that contain an In-Vision- Plus Needleless IV Connector, Model: code RYM5001 InVision-Plus Needless IV Connectors are used for single patient use in IV and blood administration sets without the need for needles, thus eliminating the potential for needle-stick injuries during use.

Why the product was recalled

Convenience Kits contain an In-Vision-Plus Needleless IV Connector, code RYM5001 that is being recalled by the manufacturer, RyMed Technologies, LLC. A component molding abnormality may cause female luer threads to crack, leading to potential for the mating tubing set to become detached and for the administration to be interrupted. Other components within these kits are not affected.

Distribution information

US Distribution in states of: TX, FL, VA, TN, PA, IL, GA, and CA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.