Recall record at a glance
| Official record | Z-2586-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Medline Industries, LP |
| Report date | Jul 1, 2026 |
| Recall initiation date | Feb 27, 2026 |
| Quantity | 113,843 kits |
| Location | Northfield, IL, United States |
Official product description
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAPHY PACK-LF DYNJ0854485S DEEP BRAIN STIMULATION PACK DYNJ0842793G NEURO ACCESSORY PACK DYNJ55385A NEURO ANGIO PACK DYNJ51697B DYNJ56509G DYNJ56509I DYNJ68063 DYNJ68845 DYNJ68845A DYNJ68845B NEURO CERVICAL PACK DYNJ0101287I DYNJ0101287L NEURO PACK DYNJ82103 DYNJ82103A NEURO TRAY DYNJ67075 DYNJ67075A DYNJ67075B NEURO VASCULAR PACK DYNJ62960B NEURO VP SHUNT PACK-LF DYNJ0843121J DYNJ0843121K NEUROLOGICAL CATH PACK DYNJ63434A DYNJ63434C NEURORADIOLOGY PK DYNJ40420D RADIOLOGY ANGIO MAJOR PACK-LF DYNJ62858A RADIOLOGY NEURO DRAPE PAC DYNJ61449 RRMC ANGIO PACK DYNJ36338B
Why the product was recalled
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
Distribution information
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.