Recall record at a glance
| Official record | Z-1890-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | ConvaTec, Inc. |
| Report date | May 23, 2018 |
| Recall initiation date | Apr 25, 2018 |
| Quantity | 200 boxes of 10 |
| Location | Greensboro, NC, United States |
Official product description
Convatec Esteem(TM) Synergy Stomahesive(R), 7/8 in. / 22mm, REF 405475 Product Usage: Esteem Synergy Stomahesive with Flexible Collar 22mm Precut Skin Barrier is used together with Esteem Synergy pouches as an ostomy system intended for the management of stomal output.
Why the product was recalled
It was discovered that the stoma hole of one lots of the Esteem synergy Stomahesive Skin Barrier wafers is larger that the 7/8 inch / 22mm specified on the box label.
Distribution information
US Nationwide Distribution
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.