B. Braun Medical CONTIPLEX Continuous Peripheral Nerve Block Tray, CONTIPLEX Stirn recall — FDA Z-1576-2015

There is a potential for a hole to be present in the outer tray of a limited number of pain control trays. This issue may cause the integrity of the…

Class IIclassification
Z-1576-2015official record ID
May 13, 2015report date

Recall record at a glance

Official recordZ-1576-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmB. Braun Medical, Inc.
Report dateMay 13, 2015
Recall initiation dateMar 27, 2015
Quantity5000
LocationAllentown, PA, United States

Official product description

CONTIPLEX Continuous Peripheral Nerve Block Tray, CONTIPLEX Stirn Continuous Nerve Block Tray, Custom Epidural and Anesthesia Tray, Custom Epidural Anesthesia Tray, Custom Kit, Custom Single Dose Epidural Anesthesia Tray, Custom Spinal Anesthesia Tray, Custom Spinal Epidural Anesthesia Tray, ESPOCAN Combined Spinal and Epidural Anesthesia Tray, PERIFIX Continuous Epidural Anesthesia Tray, PERIFIX FX Continuous Epidural Tray, PERIFIX One Continuous Epidural Pediatric Tray; Utilization of Product: Injection of local anesthetics to provide regional anesthesia for pain management. Route of administration is injection into subarachnoid space or epidural space or tissue surrounding peripheral nerve bundle(s). Custom Epidural and Anesthesia Tray

Why the product was recalled

There is a potential for a hole to be present in the outer tray of a limited number of pain control trays. This issue may cause the integrity of the sterile barrier to be compromised. A compromised sterile barrier poses a risk of inflammatory response and/or possible contamination of the spinal fluid or epidural space.

Distribution information

Nationwide Distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.