FDA recall record

The Metrix Company Container fill and transfer sets, 250 mL and 500 mL recall — FDA Z-2264-2016

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing…

Class IIclassification
Z-2264-2016official record ID
Aug 3, 2016report date

Recall record at a glance

Official recordZ-2264-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmThe Metrix Company
Report dateAug 3, 2016
Recall initiation dateJun 23, 2016
Quantity1,350 cases
LocationDubuque, IA, United States

Official product description

Container fill and transfer sets, 250 mL and 500 mL, . .

Why the product was recalled

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

Distribution information

Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.