FDA recall record

Biochemical Diagnostics CONSULT(TM) diagnostics hCG CONTROLS Kit, Part/Reorder Number 5011 recall — FDA Z-0328-2016

CONSULT diagnostics hCG Controls have discrepant storage temperatures listed on the kit labels when compared with the individual vial labels and the…

Class IIIclassification
Z-0328-2016official record ID
Dec 2, 2015report date

Recall record at a glance

Official recordZ-0328-2016
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmBiochemical Diagnostics Inc.
Report dateDec 2, 2015
Recall initiation dateSep 14, 2015
QuantityDomestic: 4,473 kits
LocationEdgewood, NY, United States

Official product description

CONSULT(TM) diagnostics hCG CONTROLS Kit, Part/Reorder Number 5011, IVD Product Usage - CONSULT Diagnostics hCG Controls are intended to validate the performance of qualitative hCG urine procedures and immunochromatographic devices.

Why the product was recalled

CONSULT diagnostics hCG Controls have discrepant storage temperatures listed on the kit labels when compared with the individual vial labels and the package insert.

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.