FDA recall record

MicroPort Orthopedics Conserve Cups: REF 38BF3642, REF 38BF3642SP, REF 38BF3844, REF recall — FDA Z-0688-2017

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing…

Class IIIclassification
Z-0688-2017official record ID
Dec 7, 2016report date

Recall record at a glance

Official recordZ-0688-2017
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmMicroPort Orthopedics Inc.
Report dateDec 7, 2016
Recall initiation dateJun 21, 2016
Quantity68,221 units
LocationArlington, TN, United States

Official product description

Conserve Cups: REF 38BF3642, REF 38BF3642SP, REF 38BF3844, REF 38BF3844SP, REF 38BF4046, REF 38BF4046SP, REF 38BF4248, REF 38BF4248SP, REF 38BF4450, REF 38BF4450SP, REF 38BF4652, REF 38BF4652SP, REF 38BF4854, REF 38BF5056, REF 38BF5258, REF 38BF5258SP, REF 38BF5460, REF 38BF5460SP, REF 38BF5662, REF 38BF5662SP, REF 38BF5864, REF 38BF5864SP, REF 38BF6066, REF 38BF6066SP, REF 38QH3644, REF 38QH3846, REF 38QH4048, REF 38QH4250, REF 38QH4452, REF 38QH4654, REF 38QH4856, REF 38QH5058, REF 38QH5260, REF 38QH5462, REF 38QH5664, REF 38QH5666, REF 38SC3650, REF 38SC3852, REF 38SC4054, REF 38SC4256, REF 38SC4458, REF 38SC4660, REF 38SC4862, REF 38SC5064, REF 38SC5266, REF 38SC5468, REF 38SC5670, REF 38SC5672 For use with the conserve metal head

Why the product was recalled

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Distribution information

Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.