Recall record at a glance
| Official record | Z-1323-2015 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | ArthroCare Medical Corporation |
| Report date | Apr 1, 2015 |
| Recall initiation date | Feb 20, 2015 |
| Quantity | 15 units |
| Location | Austin, TX, United States |
Official product description
Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit; a reusable, non-sterile flow control cable; a single use, disposable, sterile Wand. The Coblator II is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery.
Why the product was recalled
The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).
Distribution information
US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.