FDA recall record

ArthroCare Medical Coblator II System, 120V recall — FDA Z-1323-2015

The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the…

Class IIclassification
Z-1323-2015official record ID
Apr 1, 2015report date

Recall record at a glance

Official recordZ-1323-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmArthroCare Medical Corporation
Report dateApr 1, 2015
Recall initiation dateFeb 20, 2015
Quantity15 units
LocationAustin, TX, United States

Official product description

Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit; a reusable, non-sterile flow control cable; a single use, disposable, sterile Wand. The Coblator II is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery.

Why the product was recalled

The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).

Distribution information

US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records