FDA recall record

Roche Diagnostics Operations cobas p 512 pre-analytical system Pre-analytical sample recall — FDA Z-1233-2016

Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT)…

Class IIclassification
Z-1233-2016official record ID
Mar 30, 2016report date

Recall record at a glance

Official recordZ-1233-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRoche Diagnostics Operations, Inc.
Report dateMar 30, 2016
Recall initiation dateFeb 23, 2016
Quantity35 Units
LocationIndianapolis, IN, United States

Official product description

cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.

Why the product was recalled

Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.

Distribution information

US Distribution including Puerto Rico and to the states of :TX, OH, TN, AZ, WA, MI and GA

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.