FDA recall record

Roche Diagnostics Operations cobas infinity central lab running software version 3.01.03 recall — FDA Z-2444-2021

Under specific circumstances created by the user, the cobas e flow test results could be replaced by an automatic result sent to the lab LIS. This…

Class IIclassification
Z-2444-2021official record ID
Sep 22, 2021report date

Recall record at a glance

Official recordZ-2444-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRoche Diagnostics Operations, Inc.
Report dateSep 22, 2021
Recall initiation dateJun 24, 2021
Quantity2058 licenses worldwide, 79 licenses in US
LocationIndianapolis, IN, United States

Official product description

cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001

Why the product was recalled

Under specific circumstances created by the user, the cobas e flow test results could be replaced by an automatic result sent to the lab LIS. This automatic result could be released and interpreted by your Laboratory Information System (LIS) as a false positive or negative result. It should be noted that the "correct" result would still be available on the instrument.

Distribution information

US Nationwide distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.