FDA recall record

Roche Molecular Systems cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2 recall — FDA Z-2254-2016

During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation…

Class IIclassification
Z-2254-2016official record ID
Jul 27, 2016report date

Recall record at a glance

Official recordZ-2254-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRoche Molecular Systems, Inc.
Report dateJul 27, 2016
Recall initiation dateApr 5, 2016
Quantity735 devices EGFR; 509 devices cfDNA
LocationBranchburg, NJ, United States

Official product description

cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 07247737190 Device Identifier: EGFR v2.0: 00875197005448 cfDNA: 00875197005424 Product Usage: A real time PCR Test for the qualitative detection and identifications of mutations n exons 18, 19, 20 and 21 of the epidermal growth factor receptor (EGFR) gene in DNA derived from formalin-fixed paraffin-embedded (FFPET) tumor tissue or plasma from non-small cell lung cancer (NSCLC) patients. The test also is intended to be used as an aid in selecting patients with NSCLC for therapy with an EGFR tyrosine kinase inhibitor (TKI).

Why the product was recalled

During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) result with the cobas EGFR mutation Text V2.0 when utilizing the cobas cfDNA Sample Preparation kit.

Distribution information

Foriegn Distribution Austria, Belgium, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hungary , Italy, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Switzerland, Turkey, United Kingdom cfDNA: Austria, Belgium, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hungary , Italy, Lithuania, Luxembourg, Netherlands, Poland, Romania, Slovakia, Slovenia, Spain, Switzerland, Turkey, United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records