Recall record at a glance
| Official record | Z-2223-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Roche Diagnostics Corporation |
| Report date | Jun 14, 2017 |
| Recall initiation date | May 9, 2017 |
| Quantity | 3 |
| Location | Indianapolis, IN, United States |
Official product description
Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system
Why the product was recalled
a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.
Distribution information
US distribution to WA only.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.