FDA recall record

Roche Diagnostics Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a recall — FDA Z-2223-2017

a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can…

Class IIclassification
Z-2223-2017official record ID
Jun 14, 2017report date

Recall record at a glance

Official recordZ-2223-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRoche Diagnostics Corporation
Report dateJun 14, 2017
Recall initiation dateMay 9, 2017
Quantity3
LocationIndianapolis, IN, United States

Official product description

Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system

Why the product was recalled

a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.

Distribution information

US distribution to WA only.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.