Recall record at a glance
| Official record | Z-0689-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class III |
| FDA record status | Terminated |
| Recalling firm | MicroPort Orthopedics Inc. |
| Report date | Dec 7, 2016 |
| Recall initiation date | Jun 21, 2016 |
| Quantity | 108,362 units |
| Location | Arlington, TN, United States |
Official product description
Cobalt Chromium Heads Transcend: 26000017, 26000018, 26000019, 26000020, 26000021, 26000022, 26000023, 26000024, 26000025, 26000026, 26000027, 26000028 Conserve: 38013800, 38013804, 38013835, 38014000, 38014004, 38014035, 38014200, 38014204, 38014235, 38014400, 38014404, 38014435, 38014600, 38014604, 38014635, 38014800, 38014804, 38014835, 38015000, 38015004, 38015035, 38015200, 38015204, 38015235, 38015400, 38015404, 38015435, 38015600, 38015604, 38015635, 38AC3600, 38AC3800, 38AC4000, 38AC4200, 38AC4400, 38AC4600, 38AC4800, 38AC5000, 38AC5200, 38AC5400, 38AC5600, 38AM3600, 38AM3604, 38AM3635, 38AM3800, 38AM3804, 38AM3835, 38AM4000, 38AM4004, 38AM4035, 38AM4200, 38AM4204, 38AM4235, 38AM4400, 38AM4404, 38AM4435, 38AM4600, 38AM4604, 38AM4635, 38AM4800, 38AM4804, 38AM4835, 38AM5000, 38AM5004, 38AM5035, 38AM5200, 38AM5204, 38AM5235, 38AM5400, 38AM5404, 38AM5435, 38AM5600, 38AM5604, 38AM5635, PL38HM4400, PL38HM4404, PL38HM4435, PL38HM4600, PL38HM4604, PL38HM4635, PL38HM4800, PL38HM4804, PL38HM4835, PL38HM5000, PL38HM5004, PL38HM5035, PL38HM5200, PL38HM5204, PL38HM5235
Why the product was recalled
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Distribution information
Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.
How to read the classification
FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.