FDA recall record

Ethicon Coated VICRYL RAPIDETM (Polyglactin 910) Suture recall — FDA Z-0693-2018

On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910)…

Class IIclassification
Z-0693-2018official record ID
Mar 7, 2018report date

Recall record at a glance

Official recordZ-0693-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmEthicon, Inc.
Report dateMar 7, 2018
Recall initiation dateAug 17, 2017
Quantity129,372 Eaches (Internationally)
LocationSomerville, NJ, United States

Official product description

Coated VICRYL RAPIDETM (Polyglactin 910) Suture

Why the product was recalled

On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.

Distribution information

Nationally

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.