Recall record at a glance
| Official record | Z-0803-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Roche Diagnostics Operations, Inc. |
| Report date | Jan 13, 2021 |
| Recall initiation date | Dec 2, 2020 |
| Quantity | 212,093 units |
| Location | Indianapolis, IN, United States |
Official product description
CoaguChek XS Prof (Professional) Meters- IVD professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. Catalog: 04837975001 Units could consist of any of the following strip configurations: 6 count, 24 count, 2x 24 count (i.e. 48 strips)* *CoaguChek XS 3x48 promo LTC HH Catalog: 06298176001 CoaguChek XS 6x24 promo LTC HH Catalog:08468699001 CoaguChek XS 4x48 promo Catalog: 08468745001 CoaguChek XS 6x48 promo Catalog: 08468753001
Why the product was recalled
Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
Distribution information
US Nationwide
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.