Clinical Innovations Clearview Total Uterine Manipulator (model UM750), The device is recall — FDA Z-2118-2020

An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or…

Class IIclassification
Z-2118-2020official record ID
Jun 3, 2020report date

Recall record at a glance

Official recordZ-2118-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmClinical Innovations, LLC
Report dateJun 3, 2020
Recall initiation dateMay 30, 2019
Quantity280 devices
LocationMurray, UT, United States

Official product description

Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICAL INNOVATIONS Manufacturer: Clinical Innovations, LLC* * *Reorder Number: UM750* * *LOT 181183* * *." Each device is accompanied by Instruction for Use which is similarly labeled.

Why the product was recalled

An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or colpotomy cup. This may result in a delay in surgery and additional surgical interventions to completed the procedure. There was a report of an injury.

Distribution information

International distribution to the country Japan.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.