FDA recall record

CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) recall — FDA Z-2888-2018

A cover failed leak testing during validation testing of new needle guides from a replacement tool.

Class IIclassification
Z-2888-2018official record ID
Aug 29, 2018report date

Recall record at a glance

Official recordZ-2888-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCivco Medical Instruments Co. Inc.
Report dateAug 29, 2018
Recall initiation dateFeb 23, 2018
Quantity177 boxes
LocationKalona, IA, United States

Official product description

CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.

Why the product was recalled

A cover failed leak testing during validation testing of new needle guides from a replacement tool.

Distribution information

Worldwide Distribution - US Nationwide in the states of GA, KY, CA, NY , NJ, TX, IL, MA, RI, AZ, and IA and countries of Ukraine, Hong Kong, Italy, Netherlands, Switzerland, United Kingdom, Denmark, Germany, Belgium, Spain, Japan, China, and Canada

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.