FDA recall record

Covidien CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be recall — FDA Z-1334-2021

The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration,…

Class IIclassification
Z-1334-2021official record ID
Apr 7, 2021report date

Recall record at a glance

Official recordZ-1334-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCovidien LLC
Report dateApr 7, 2021
Recall initiation dateFeb 26, 2021
Quantity62,551
LocationMansfield, MA, United States

Official product description

CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.

Why the product was recalled

The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.

Distribution information

Worldwide distribution - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.