FDA recall record

Cepheid recall — FDA Z-2584-2024

Specimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure, cross…

Class IIclassification
Z-2584-2024official record ID
Aug 21, 2024report date

Recall record at a glance

Official recordZ-2584-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCepheid
Report dateAug 21, 2024
Recall initiation dateJun 20, 2024
Quantity651,150
LocationSunnyvale, CA, United States

Official product description

Cepheid, Sample Collection Device, Part: 900-0370

Why the product was recalled

Specimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure, cross contamination causing false positive results for other specimens, and/or delay to test results.

Distribution information

Worldwide distribution: US (nationwide): IL, TX, OH, LA, NY, CA, MD, IA, MN, KY, NH, WI, MO, NC, UT, WA, FL, NJ, SC, CO, PA, SD, NE, MT, WV, GA, MA, RI, CT, DE, MS, OK, VA, ID, AZ, OR, MI, IN, TN, AL, KS, ME, AR, NV, HI, AK, NM, WY, VT, ND, DC OUS (foreign) to countries of: Chile, Ireland, Oman, Nicaragua, Germany, San Marino, Ecuador, New Caledonia, France, Austria, United Kingdom, Finland, Italy, Hong Kong, Kuwait, United Arab Emirates, Netherlands, Belgium, Monaco, Colombia, Spain, Portugal, Poland, Switzerland, Denmark, Jersey, Mexico

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.