FDA recall record

Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the recall — FDA Z-0120-2020

A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.

Class IIclassification
Z-0120-2020official record ID
Oct 16, 2019report date

Recall record at a glance

Official recordZ-0120-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCenturion Medical Products Corporation
Report dateOct 16, 2019
Recall initiation dateAug 28, 2019
Quantity1530 each
LocationWilliamston, MI, United States

Official product description

Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 310CR

Why the product was recalled

A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.