FDA recall record

GE Medical Systems Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a recall — FDA Z-1569-2026

There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the…

Class IIclassification
Z-1569-2026official record ID
Mar 25, 2026report date

Recall record at a glance

Official recordZ-1569-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE Medical Systems, LLC
Report dateMar 25, 2026
Recall initiation dateJan 30, 2026
Quantity1568 units
LocationWaukesha, WI, United States

Official product description

Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 5826659-027, 5826659-029, 5826659-032, 5826659-033, 5826659-028, 5826659-034, 5826659-030, 5826659-019, 5826659-022, 5826659-026, 5826659-023, 5826659-031, 5826659-025, 5826659-002, 5826659-021, 5826659-037, 5826659-014, 5826659-016, 5826659-013

Why the product was recalled

There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.

Distribution information

Worldwide - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.