FDA recall record

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION Ceiling Type X-Ray Tube Support Models CH-200 recall — FDA Z-0499-2026

Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the…

Class IIclassification
Z-0499-2026official record ID
Nov 19, 2025report date

Recall record at a glance

Official recordZ-0499-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Report dateNov 19, 2025
Recall initiation dateOct 16, 2025
Quantity1,766 devices
LocationKyoto, Japan

Official product description

Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system.

Why the product was recalled

Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.

Distribution information

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SM, SC, SD, TN, TX, VA, WA, and WI. The countries of Dominica, Grand Bahama, Haiti, and Saint Maarten.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.