FDA recall record

Biosense Webster (israel) CARTO 3 System (Software Version V7.1.80), CARTO VISITAG Module recall — FDA Z-2880-2020

Software defect may result in disapperance of tag sites during recalculation and lead to additional ablation sites. This may lead to prolongation of the…

Class IIclassification
Z-2880-2020official record ID
Sep 2, 2020report date

Recall record at a glance

Official recordZ-2880-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBiosense Webster (israel), Ltd.
Report dateSep 2, 2020
Recall initiation dateJun 22, 2020
QuantityN/A
LocationYoqne'Am Illit, N/A, Israel

Official product description

CARTO 3 System (Software Version V7.1.80), CARTO VISITAG Module - Product Usage: FGThe intended use of the system is catheter-based cardiac electrophysiological (EP) procedures. The system provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure, Product Number: -5400-00, UDI: 10846835000870

Why the product was recalled

Software defect may result in disapperance of tag sites during recalculation and lead to additional ablation sites. This may lead to prolongation of the procedure and, in extremely rare circumstances, cardiac perforation.

Distribution information

Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA, CO, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV, and WI *Three (3) U.S. government consignees. The countries of Australia, Austria, Belgium, Chile, China, Colombia, Czechia, Denmark, Finland, France, Germany, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Latvia, Netherlands, Norway, Poland, Portugal, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.