FDA recall record

Boston Scientific Carotid WALLSTENT Monorail Endoprosthesis Closed Cell Self-Expanding recall — FDA Z-2273-2025

Certain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while…

Class Iclassification
Z-2273-2025official record ID
Aug 20, 2025report date

Recall record at a glance

Official recordZ-2273-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmBoston Scientific Corporation
Report dateAug 20, 2025
Recall initiation dateJul 7, 2025
Quantity26,570 (1,333 US; 25,174 OUS)
LocationMaple Grove, MN, United States

Official product description

Carotid WALLSTENT Monorail Endoprosthesis Closed Cell Self-Expanding Stent, stent and stent delivery system. Material numbers: H7493915010240 CAROTID WALLSTENT MONORAIL 10.0-24; H7493915010310 CAROTID WALLSTENT MONORAIL 10.0-31; H749391506220 CAROTID WALLSTENT MONORAIL 6.0-22; H749391508210 CAROTID WALLSTENT MONORAIL 8.0-21; H749391508290 CAROTID WALLSTENT MONORAIL 8.0-29; H965SCH647010 CAROTID WALLSTENT MONORAIL 6.0-22; H965SCH647070 CAROTID WALLSTENT MONORAIL 8.0-21; H965SCH647080 CAROTID WALLSTENT MONORAIL 8.0-29; H965SCH647090 CAROTID WALLSTENT MONORAIL 8.0-36; H965SCH647120 CAROTID WALLSTENT MONORAIL 10.0-24; H965SCH647130 CAROTID WALLSTENT MONORAIL 10.0-31; H965SCH647140 CAROTID WALLSTENT MONORAIL 10.0-37; M001719000 CAROTID WALLSTENT MR 6X22 5F 135CM; M001719010 CAROTID WALLSTENT MR 8X21 5F 135CM; M001719020 CAROTID WALLSTENT MR 8X29 5F 135CM; M001719030 CAROTID WALLSTENT MR 8X36 5F 135CM; M001719040 CAROTID WALLSTENT MR 10X24 5.9F 135CM; M001719050 CAROTID WALLSTENT MR 10X31 5.9F 135CM; M001719060 CAROTID WALLSTENT MR 10X37 5.9F 135CM;

Why the product was recalled

Certain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the guidewire or embolic protection device (EPD) after successful stent deployment. The most serious potential adverse health consequence is stroke. Other risks include delay to procedure, vessel injury, vessel spasm, or stent disruption/damage necessitating additional intervention.

Distribution information

Domestic US Nationwide distribution including Puerto Rico. International distribution worldwide.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.