Draeger Medical Carina Ventilator, ventilator recall — FDA Z-1954-2018

The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in…

Class IIclassification
Z-1954-2018official record ID
Jun 6, 2018report date

Recall record at a glance

Official recordZ-1954-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDraeger Medical, Inc.
Report dateJun 6, 2018
Recall initiation dateApr 26, 2018
Quantity8 units
LocationTelford, PA, United States

Official product description

Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel.

Why the product was recalled

The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.

Distribution information

Distributed to the states of CT, IL, MN, NV, SD, and WA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 25, 2024

Atlan A300XL recall — FDA Z-0668-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.
Medical DeviceDec 25, 2024

Atlan A300 recall — FDA Z-0667-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.
Medical DeviceNov 20, 2024

Atlan A350XL recall — FDA Z-0305-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.
Medical DeviceNov 20, 2024

Atlan A350 recall — FDA Z-0304-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.