CareFusion EnVe and ReVel Ventilators recall — FDA Z-2593-2014

CareFusion has identified a potential for damage to the power connector on the EnVe and ReVel ventilators due to misconnection by the operator. The input…

Class Iclassification
Z-2593-2014official record ID
Oct 1, 2014report date

Recall record at a glance

Official recordZ-2593-2014
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmCareFusion 203, Inc.
Report dateOct 1, 2014
Recall initiation dateAug 8, 2014
Quantity1729
LocationMinneapolis, MN, United States

Official product description

CareFusion EnVe and ReVel Ventilators. The EnVe/ ReVel Ventilator is designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs.). It is suitable for service in hospital, transport and homecare as a source of continuous or intermittent positive pressure ventilator support, delivered invasively or non-invasively. These ventilators are restricted medical devices intended for operation by a trained personnel under the direction of a physician and in accordance with all applicable state laws and regulations.

Why the product was recalled

CareFusion has identified a potential for damage to the power connector on the EnVe and ReVel ventilators due to misconnection by the operator. The input port on the ventilators must have the adapters correctly aligned to function properly. Incorrect alignment may cause the pins in the connector to bend, causing the PTV AC power adapter and/or PTV Auto Lighter Power Cord to be shortened, and pre

Distribution information

Nationwide Distribution including AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MI, MN, MS,M0, NE, NV,NH,NH, NY NMC, OH, OR,PA, RID, SC, TN, TX, UT VT, VA, WA, and WI.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.