Recall record at a glance
| Official record | Z-1360-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Ongoing |
| Recalling firm | CareFusion 303, Inc. |
| Report date | Mar 11, 2020 |
| Recall initiation date | Feb 4, 2020 |
| Quantity | 1286898 |
| Location | San Diego, CA, United States |
Official product description
CareFusion Alaris Pump Module, Model 8100
Why the product was recalled
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders
Distribution information
Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.