FDA recall record

Datascope Cardiosave Rescue recall — FDA Z-1232-2023

Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks.…

Class IIclassification
Z-1232-2023official record ID
Mar 15, 2023report date

Recall record at a glance

Official recordZ-1232-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDatascope Corp.
Report dateMar 15, 2023
Recall initiation dateFeb 7, 2023
Quantity47 units (44 US, 3 OUS)
LocationMahwah, NJ, United States

Official product description

Cardiosave Rescue. Model Number: 0998-00-0800-83.

Why the product was recalled

Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.

Distribution information

Nationwide US distribution. Worldwide international distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 25, 2026

Datascope Battery Charging Station recall — FDA Z-1587-2026

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has…

Class IIIFDA: OngoingDatascope Corp.