FDA recall record

Datascope Cardiosave Hybrid IntraAortic Balloon Pump recall — FDA Z-0336-2022

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to…

Class Iclassification
Z-0336-2022official record ID
Dec 22, 2021report date

Recall record at a glance

Official recordZ-0336-2022
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmDatascope Corp.
Report dateDec 22, 2021
Recall initiation dateNov 15, 2021
Quantity5,566 (US); 6,354 (OUS)
LocationMahwah, NJ, United States

Official product description

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.

Why the product was recalled

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.

Distribution information

Worldwide distribution - US Nationwide.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 25, 2026

Datascope Battery Charging Station recall — FDA Z-1587-2026

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has…

Class IIIFDA: OngoingDatascope Corp.