Toshiba American Medical Systems Cardiac Function Analysis Software Product Data No recall — FDA Z-0471-2015

Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire…

Class IIclassification
Z-0471-2015official record ID
Dec 3, 2014report date

Recall record at a glance

Official recordZ-0471-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmToshiba American Medical Systems Inc.
Report dateDec 3, 2014
Recall initiation dateNov 5, 2014
Quantity1 unit within the U.S. only
LocationTustin, CA, United States

Official product description

Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301C X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits scanning over large areas at the high-speed of 0.35 s per rotation. The system features are; 160-mm-wide area detector, station, gantry and patient couch operation controls, image reconstruction, high-speed volume scan, high-quality images, selectable image slice thickness, dual-monitor system, improved image diagnosis functions, high-speed volume data workflow, exposure reduction, ECG-gated scan and reconstruction, and SUREFluoro. The CT cardiac function analysis software is used to analyze cardiac function using data obtained by performing cardiac MPR processing for ECG-gated scan data according to the cardiac axis. Aquilion ONE is a dynamic volume CT system that supports whole-body scanning. This 320-detector row system generates 640 slices per rotation using the coneXact reconstruction algorithm. In addition, the high-speed rotation mechanism and the fast reconstruction unit of the system allow the rapid image acquisition to further improve throughput in CT examinations.

Why the product was recalled

Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters.

Distribution information

US Nationwide Distribution to CA, MA, MD

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.